Rotem TelePole Gamma Survey Monitor
HARGA
Rp 89,000,000.00
= = = = = = = = = = = = = = = = = = = = = =
(UPDATE 4 FEBRUARI 2013)
= = = = = = = = = = = = = = = = = = = = = =
> Langsung kalibrasi.
> Harga sudah termasuk kalibrasi pertama kali dan pengiriman ke tempat tujuan.
> Jika menghendaki transaksi an. Perusahaan, maka harga masih belum termasuk ppn 10%. (Jika ingin transaksi resmi bukan pribadi).
The TelePole is a gamma survey meter, with the detector mounted on an extendable pole designed for monitoring less accessible areas or providing " safe distance" operation. The four segment telescopic pole extends to 337cm ( 11 feet) and collapses to 107cm ( 3.5 feet) with the detector mounted at the pole end. There are two detector configurations are available. The WR ( wide range) provides measurements over the range 0.5µ Sv/ hr to 10Sv/ hr ( 50µ R/ hr - 1000R/ hr) , while the HR ( high range) covers 50µ Sv/ hr to 10Sv/ hr ( 5mR/ hr - 1000R/ hr) . An optional internal detector is also available. TelePole features a 380 record data memory that can be downloaded to a PC using the new RMV Software Package. An optional barcode reader is also available which makes TelePole fully compatible with the SMARTS radiation survey mapping and tracking system.
* Rugged, lightweight construction (~2Kg)* Digilog, continuous ranging, combined digital and analog, LCD display* External detectors for either wide range or high range applications* Optional internal detector* Readout of dose rate and accumulated dose* Units of either mR or µSv with dead time correction.* Programmable alarm thresholds* Data memory for 380 records (1550 with optional extended memory)* Compatible with RMV software for pc data downloads* Smart software provides fast, yet stable response* Alarms for detector overflow, low battery and detector failure* Automatic self diagnostic routines* External headphone connection.
Display:
Large, three digit and two decade backlit LCD readout.
Continuous ranging, for display of accumulated dose, dose rate, alarm threshold value, overflow, low battery and detector failure
Measuring Range ( WR) :
0.5µ Sv/ hr to 10Sv/ hr ( 50µ R/ hr - 1000R/ hr)
Measuring Range ( HR) :
50µ Sv/ hr to 10Sv/ hr ( 5mR/ hr - 1000R/ hr)
Measuring Range:
Internal Detector Option:
0.5µ Sv/ hr to 999mSv/ hr ( 0.05mR/ hr to 99R/ hr) .
Mounted within electronics control unit.
Sensitivity:
0.3 cps/ mR/ hr ( 137Cs)
Accuracy:
± 10% of reading within measuring range
Data Logging:
380 Records ( up to 1550 records with extended memory)
Units:
mR and mR/ hr or µ Sv and µ Sv/ hr
Controls:
Four key keypad with positive feedback. Key combinations
for programming operating parameters.
Power Source:
Two 1.5V alkaline ' C' cells - provides approximately 70 hours
continuous use ( without backlighting) .
Automatic battery check under full load.
Temperature Range:
Operation: -10oC to + 50oC ( 15oF to 122oF)
Storage: -20oC to + 60oC ( -5oF to 140oF)
Humidity Range:
40% to 95% RH ( non condensing)
Case:
Black anodized aluminum
Dimensions:
Meter: 93mmW x 148mmL x 56mmH ( 3.7" x 5.8" x 2.2" )
Probe Head: 150mmL x 25mmD ( 5.9" x 1" )
Pole Length: 337cm ( 11ft) extended, 107cm ( 3.5ft) collapsed
Weight:
Meter: 680g ( 1.5 lbs) , Probe: 110g ( 0.25lbs) , Pole: 1.2Kg ( 2.7lbs)