We concern to health and safety

------------------

Selasa, 17 Juni 2014

DMC 3000 Electronic Personal Dosimeter, X-ray & Gamma


Manufacturer: Mirion Technologies (MGP)
Product ID : MG-153503

Overview

The DMC 3000 Personal Electronic Dosimeter has a rugged, high impact polycarbonate-ABS case, is small, lightweight, and is cost effective personal radiation monitoring device. They are designed to be worn on the body and keep a live record of both dose and dose rate. The DMC 3000 Personal Electronic Dosimeter is simple to use with 2 push buttons for an easy customized data and
parameters display. The DMC 3000 has a LDC display with high-quality white backlighting.

The DMC 3000 - Personal Electronic Dosimeter - X-ray & Gamma
The DMC 3000 covers a wide range of X-Ray and Gamma radiation detection field from 15keV to 7Mev and linear response to dose rate fields from natural background up to more than 10 Sv/h.  The pass-by data exchange feature gives unequal operational flexibility. In-motion reading allows dose management by sub-zone as well as real-time location tracking of personnel.


FEATURES
  • Loud alarm, 85 dB (A) typical, (> 90 dB (C) peak)
  • Vibrating alarm
  • Highly visible backlit display
  • Optional telemetry module
  • Factory upgradeable firmware
  • Simple 2-button navigation
  • Extended dose rate alarms
  • Dual ultrabright LED alarm
  • Superior X-ray and gamma energy response
  • Meets or exceeds applicable IEC and ANSI standards
  • Designed for ruggedness and durability

TECHNICAL SPECIFICATIONS:
Radiological Characteristics
  • Compliant with IEC 61526Ed. 3, ANSI 42.20(*) (*) isotropy 241Am and 137Cs with ± 75° angle Measurement and display
  • Energy response better than ± 20% (typically ± 10%)
  • Accuracy Hp(10): ≤ ± 10%* (137Cs, ~ 24 mSv/h); ≤ ± 15%*, (241Am, ~ 23 mSv/h); ≤ ± 19%** X-ray 16keV (*including ± 5% extended uncertainty k=2); (** including ± 9% extended uncertainty k=2)
  • Display dose: 1 µSv to 10 Sv (0.1 mrem to 1000 rem)
  • Display rate: 10 µSv/h to 10 Sv/h (1 mrem/h to 1000 rem/h) or 1 µSv/h to 10 Sv/h (0.1 mrem/h to 1000 rem/h) extended option. 
  • Dose Rate Linearity: < ± 20% up to 10 Sv/h (1000 rem/h) (Co and X H30 20 keV); < ± 20% up to 6 Sv/h (600 rem/h) (Pulsed X-rays 20 ms width, 1, 10 & 20 pps)
Electrical Characteristics
  • Standard AAA (LR03) 1.5 V Alkaline battery
  • 9 calendar month battery life (typical, 8 h per day,
     
    5 days per week in run mode, without excessive alarms)* 
  • 2500 h battery life in continuous run, without excessive alarm
Mechanical Characteristics
  • Rugged, high impact polycarbonate-ABS case
  • Dimensions: 86 x 56 x 21 mm (3.4 x 2.2 x 0.8 in) without clip 
  • Weight with alkaline battery and clip: < 88 g (3.1 oz) 
  • Worn by a replaceable clips (2 different style back clips or one a front-facing clip for DMC worn inside the pocket)
Environmental Characteristics
  • Temperature range: -10°C to 50°C (14°F to 122°F): deviation in response less than ± 5% 
  • Relative humidity: < 90% at 42°C (108°F) 
  • Storage: -20°C to 71°C (-4°F to 160°F) 
  • Shock, vibration and drop resistant (1.5 meter on concrete)
  • Waterproof IP67: 1m (39 in) during 1 hour 
  • EMC: complies and exceeds standards by a large margin ( compliant, certificate number: 153720)
  • MIL STD 461F RS103 (pulsed electric field): exceeds 200 V/m from 30 kHz to 5 GHz
  • MIL STD 461F RS101 (magnetic field 30 Hz - 100 kHz)
  • Factory calibration approved under ISO/IEC 17025, COFRAC accreditation Nber 2-1663

0 komentar: